Main Article Content
Abstract
Objectives: The aim of this research was to examine the average glandular dose (AGD) of radiation among different breast compositions of glandular and adipose tissue with auto-modes of exposure factor selection in digital breast tomosynthesis. Methods: This experimental study was carried out in the National Cancer Society, Kuala Lumpur, Malaysia, between February 2012 and February 2013 using a tomosynthesis digital mammography X-ray machine. The entrance surface air kerma and the half-value layer were determined using a 100H thermoluminescent dosimeter on 50% glandular and 50% adipose tissue (50/50) and 20% glandular and 80% adipose tissue (20/80) commercially available breast phantoms (Computerized Imaging Reference Systems, Inc., Norfolk, Virginia, USA) with auto-time, auto-filter and auto-kilovolt modes. Results: The lowest AGD for the 20/80 phantom with auto-time was 2.28 milliGray (mGy) for two dimension (2D) and 2.48 mGy for three dimensional (3D) images. The lowest AGD for the 50/50 phantom with auto-time was 0.97 mGy for 2D and 1.0 mGy for 3D. Conclusion: The AGD values for both phantoms were lower against a high kilovolt peak and the use of auto-filter mode was more practical for quick acquisition while limiting the probability of operator error.
Keywords
Thermoluminescent Dosimetry
Breast
Mammography
Radiation Dosage.
Article Details
How to Cite
Kamal, I., Chelliah, K. K., & Mustafa, N. (2015). Estimates of Average Glandular Dose with Auto-modes of X-ray Exposures in Digital Breast Tomosynthesis. Sultan Qaboos University Medical Journal, 15(2), 292–296. Retrieved from https://journals.squ.edu.om/index.php/squmj/article/view/2029